    Article Contents
    HU Shenghong, CHEN Aifang, LIN Shoulin, YUAN Honglin, GAO Shan, 2000. ICP-MS ANALYTICAL RESEARCH INTO 40 TRACE AND ULTRA-TRACE ELEMENTS IN GEOLOGICAL SAMPLES. Earth Science, 25(2): 186-190.
    Citation: HU Shenghong, CHEN Aifang, LIN Shoulin, YUAN Honglin, GAO Shan, 2000. ICP-MS ANALYTICAL RESEARCH INTO 40 TRACE AND ULTRA-TRACE ELEMENTS IN GEOLOGICAL SAMPLES. Earth Science, 25(2): 186-190.

    ICP-MS ANALYTICAL RESEARCH INTO 40 TRACE AND ULTRA-TRACE ELEMENTS IN GEOLOGICAL SAMPLES

    • Received Date: 1999-09-07
    • Publish Date: 2000-03-25
    • The analytical method of major, trace and ultra-trace elements is applied to geological samples using ICP-MS. In order to minimize matrix effects of inter-elements, the matrix-matched calibration solutions were prepared, simulating the composition of natural rocksobtained from statistically-calculated average values of 20 geological samples. The resulting values are then normalized. Two separate internal standards of 115 In-103Rh are selected to compensate the drift of analytical signals. Polyatomic ion interference is calibrated by measuring the oxides(MO+)and hydroxides(MOH+)with individual element solution for the ratio of MO+/M+ and MOH+/M+ and the concentrations are then equivalently calculated. The proposed method is applied to the analysis of seven standard reference materials. For the most trace elements, the relative errors between this work and the recommend values are greater than 10% and the relative standard deviations are greater than 5%.

       

    • loading
    • [1]
      Jenner G A, Longerich H P, Fryer B J, et al. Inductively coupled plasma_mass spectrometric analysis of geological samples: a critical evaluation based on case studies[J]. Chem Geol, 1990, 83: 105~118. doi: 10.1016/0009-2541(90)90143-U
      [2]
      Xie Q, Jain J, Sun M, et al. ICPMS analysis of basalt BIR_1 for trace elements[J]. Geostand Newslett, 1994, 18: 53~63. doi: 10.1111/j.1751-908X.1994.tb00504.x
      [3]
      Thompson J J, Houk R S. A study of internal standardisation in inductively coupled plasma mass spectrometry[J]. Appl Spectrosc, 1987, 41: 801~806. doi: 10.1366/0003702874448265
      [4]
      Doherty W. An internal standardization procedure for the determination of yttrium and the rare earth elements in geological materials by inductively coupled plasma_mass spectrometry[J]. Spectrochim Acta, 1989, 44B: 263~280.
      [5]
      Cheatham M M, Sangrey W F, White W M. Sources of error in external calibration ICPMS analysis of geological samples and an improved non_liner drift correction procedure[J]. Spectrochim Acta, 1993, 48B: 487~506.
      [6]
      Schonberg G. Simultaneous determination of thirty_seven trace elements in twenty_eight international rock standardsby ICPMS[J]. Geostandards Newsletter, 1993, 17: 81~97. doi: 10.1111/j.1751-908X.1993.tb00122.x
      [7]
      Eggin S M, Woodhead J D, Kinslet L, et al. Sample method for the simultaneous and precise analysis of 40 or more trace elements in geological samples by ICP_MS using enriched isotope internal standardization[J]. Chem Geol, 1996, 134: 311~326.
      [8]
      刘颖, 刘海臣, 李献华. 用ICP-MS准确测定岩石样品中的40余种微量元素[J]. 地球化学, 1996, 25: 532~558. https://www.cnki.com.cn/Article/CJFDTOTAL-DQHX606.003.htm
      [9]
      Govindaraju K. 1994 compilation of working values and sample description for 383 geostandards[J]. Geostandards Newsletter, 1994, 18(special issue): 1~158.